This test measures the time required for a waveform envelope to settle within the tolerance of its final amplitude after a sharp amplitude transition provided by the Alternate Timeslot Power feature (Option UNA). Settling time is the result of heating and cooling effects in the step attenuator and ALC loop. The rise and fall response times of the ALC loop and each attenuator segment is characterized and combined to construct the settling curve for any arbitrary transition.
|
Instrument |
Critical Specifications for This Test |
Recommended Model |
|
Spectrum Analyzer |
Marker Accuracy; Log Fidelity |
E4440A, E4443A, E4445A, E4446A, E4448A |
|
10 dB Attenuator |
none |
8491A Option 010 |
Connections and Setup Procedures
|
|
All test equipment requires a 30 minute warm-up period to ensure accurate
performance. |
Connect all test equipment as shown.
Connect GPIB cables to all GPIB-controlled test equipment.
While performing this test, follow all instructions on the controller display.