The AM accuracy is verified directly with a measurement receiver. For frequencies > 1300 MHz, the RF output from the signal generator is mixed with the output from another signal generator (local oscillator) and downconverted to produce a suitable input frequency for the measuring receiver.
|
Instrument |
Critical Specifications for This Test |
Recommended Model |
|
Measuring Receiver |
AM Demodulated Distortion (<50% AM) ≤ 0.3% |
8902A |
|
Mixer Not used for ESGs with frequency range <2 GHz |
RF/LO Freq. Range: 1 to 2800 MHz IF Frequency Range: 1.5 MHz RF 1 dB compression: +8 dBm RF/LO Freq. Range: 500 to 9000 MHz IF Frequency Range: 50 MHz |
M/A-COM MDC-174 M/A-COM MDC-164 |
|
Local Oscillator Not used for ESGs with frequency range <2 GHz |
Frequency Range: 500 to 2500 MHz Amplitude: 16 dBm |
E8241A |
|
|
||
Connections and Setup Procedures
|
|
All test equipment requires a 30 minute warm-up period to ensure accurate
performance. |
Connect all test equipment as shown.
Connect GPIB cables to all GPIB-controlled test equipment.
While
performing this test, follow all instructions on the controller display.
Measuring Receiver Calibration Setup

AM Accuracy CW Frequency for <= 1300 MHz and Test Setup

AM Accuracy CW Frequency > 1300 MHz