The signal generator is configured for an external burst input. The function generator output is switched between 0 and &endash;1.05 Vdc to switch the ESG burst modulation on and off. A spectrum analyzer is used to measure the power of the fundamental (CW) frequency of the ESG with the burst on and off. The difference in these power levels is the Burst ON/OFF ratio, in dB.
|
Instrument |
Critical Specifications for This Test |
Recommended Model |
|
Function Generator |
none |
33120A, 33250A |
|
Spectrum Analyzer |
Marker Accuracy; Log Fidelity |
E4440A, E4443A, E4445A, E4446A, E4448A, |
|
Digital Voltmeter |
none |
3458A |
Connections and Setup Procedures
|
|
All test equipment requires a 30 minute warm-up period to ensure accurate
performance. |
Connect all test equipment as shown.
Connect GPIB cables to all GPIB-controlled test equipment.
While
performing this test, follow all instructions on the controller display.