DC FM Carrier Offset is a measure of carrier-frequency shift due to residual DC on the ESG’s EXT1 DC -coupled path when FM mode is selected. Any DC level in this path will cause a constant offset (deviation) in the CW frequency.
For this test, the ESG is set to a selected carrier frequency and FM is enabled. The EXT 1 input is selected and left open. A specific value of FM deviation is selected and a DC FM calibration is executed.
For FM ON and FM OFF, the carrier frequency is measured with a frequency counter. Any shift in the carrier frequency must be due to a residual DC voltage somewhere in the EXT 1 input path.

|
Instrument |
Critical Specifications for This Test |
Recommended Model |
|
Frequency Counter |
Frequency Resolution ≥ 0.1 Hz at 1 GHz |
HP 53131A or HP 53132A, |
Connections and Setup Procedures
|
|
All test equipment requires a 30 minute warm-up period to ensure accurate
performance. |
Connect all test equipment as shown.
Connect GPIB cables to all GPIB-controlled test equipment.
While performing this test, follow all instructions on the controller display.