FM distortion is defined as the square root of the ratio of the power contained in the harmonics of a demodulated FM waveform to the power of the fundamental component of the demodulated waveform.

This is essentially a total harmonic distortion (THD) measurement and is made directly by the 8903B Audio Analyzer. The THD indicated by the 8903B includes the distortion added by the measuring receiver in demodulating the FM signal form the DUT.
|
Instrument |
Critical Specifications for This Test |
Recommended Model |
|
Measuring Receiver |
FM Demodulated Distortion < 0.1% |
8902A |
|
Audio Analyzer |
Distortion Accuracy |
8903B |
Connections and Setup Procedures
|
|
All test equipment requires a 30 minute warm-up period to ensure accurate performance. |
Connect all test equipment as shown.
Connect GPIB cables to all GPIB-controlled test equipment.
While
performing this test, follow all instructions on the controller display.
Measuring Receiver Calibration Setup

FM Distortion Test Setup