A dc voltage is used to switch the DUT pulse modulation ON and OFF. A spectrum analyzer is used to measure the power of the fundamental (CW) frequency of the DUT with the pulse on and off. The difference in these power levels is the Pulse ON/OFF ratio, in dB
|
Instrument |
Critical Specifications for This Test |
Recommended Model |
|
Function Generator |
none |
33120A, 33250A |
|
Spectrum Analyzer |
Marker Accuracy; Log Fidelity |
E4440A, E4443A, E4445A, E4446A, E4448A, |
Connections and Setup Procedures
|
|
All test equipment requires a 30 minute warm-up period to ensure accurate
performance. When testing Option 1E6, use the ESG PULSE IN connector in place of the EXT 2 INPUT. |
Connect all test equipment as shown.
Connect GPIB cables to all GPIB-controlled test equipment.
While
performing this test, follow all instructions on the controller display.

Option 1E6 Test Setup